This series of STERICOX fully automatic autoclave comes with most advanced features that are required by highly precise sterilization process. GR series vertical autoclave features fully automatic operation, lid interlock, several fixed cycles and steam collection bottle etc. These autoclaves are fitted with microprocessor based control panel with fixed cycles including Liquid, Liquid with warming, Solid, wrapped instrument, Fabric, Rubber, Fast, Waste and Agar Melting and one self-defined cycle. Door safety provision is in-built feature. Lid close checking function prevents cycling running before lid properly closed. Standard models are available in 60, 85 and 110 Liters.
Technical Specifications
Model |
STXGR60DA |
STXGR85DA |
STXGR110DA |
Model type |
Fully automatic |
Fully automatic |
Fully automatic |
Volume |
60 Liters |
85 Liters |
110 Liters |
Overall dimensions (W*D*H) |
660*644*980 mm |
660*644*980 mm |
660*644*1180 mm |
Inner chamber dimensions (Dia.*L) |
400*505 mm |
400*700 mm |
400*895 mm |
Power |
2900W |
4600W |
4600W |
Basket type |
Stainless steel baskets |
Basket nos. |
2 nos. of 360*200 mm |
2 nos. of 360*280 mm |
3 nos. of 360*250 mm |
Power supply |
220V/50Hz Single Phase 32A |
Temperature accuracy |
0.1°C |
Working environment |
5-45°C, relative humidity 10% to 85% |
Inner chamber material |
Stainless steel 304 |
Sterilization temperature |
105°C to 138°C |
Sterilization time range |
1-6000 minutes |
Melting temperature |
60°C to 100°C |
Warming temperature |
46°C to 60°C |
Warming time range |
1-9999 minutes |
Cooling lock OPEN temperature |
40°C-80°C for liquid cycle, 40-99°C for other cycles |
Sterilization cycle |
Liquid, Liquid with warming, Solid, Wrapped instrument, Fabric, Rubber, Fast, Waste, Agar Melting, Self-defined |
Memory |
20 cycles |
Exhaust mode |
Automatic internal discharging with a steam collecting bottle inside. Exhaust speed can be selected by 6 levels |
Lid opening way |
Flip open type |
Steam collector and water tank |
Embedded 2 steam collectors |
Present range of timer |
Can be preset and auto start up to 6 days in advance |
Display range of pressure gauge |
-0.1 to 0.5Mpa |
Pressure |
Max working pressure 0.27 Mpa design pressure 0.35Mpa |
Controller |
“Inspiration” fast speed microcomputer controller |
Safety device |
Eight column evenly distributed interlocking device, electric double inner locks, lid closing checking, dry scorch protection, overpressure protection, safety valve, water level sensor, over temperature protection, temperature monitor, over current and short circuit protection, leakage protection, cooling lock, anti-scald chamber cover and bench, automatic trouble shooting |